Main features:
Digital display resolution: 0.01 ¡æ
Temperature fluctuation value: ¡À 0.05 ¡æ
Temperature uniformity: ¡À 0.05 ¡æ horizontal/vertical
Compact structure, lowered center of gravity, strong stability
Technical features:
The fully enclosed environmentally friendly compressor refrigeration system uses environmentally friendly refrigerants that comply with the Montreal Convention.
¡ñ Temperature range: -120~100 ¡æ can be selected (the medium can be pure water, ethylene glycol, alcohol, etc. according to the temperature range);
¡ñ Temperature control function: Equipped with an integrated next-generation PID automatic temperature control program, with a built-in high-precision PT100 temperature sensor rated AAA. Equipped with a large screen high brightness LCD screen, displaying in both Chinese and English, with a digital resolution of 0.01 ¡æ, ensuring precise and stable temperature control.
Capacity range: 6L, 10L, 15L, 20L, 30L, 50L, 100L, and non-standard design and production can also be fully customized according to user requirements.
Internal circulation function: Adopting a non heat centrifugal circulation pump to avoid the influence of traditional water pumps on the tank temperature field due to their own heat generation. The internal circulation system is designed based on the principle of porous U-shaped reflux circulation, achieving turbulent and highly uniform flow and temperature fields within the tank.
The body shell adopts anti-aging electrostatic spraying technology, and the inner liner and tabletop are made of high-quality S304 brushed stainless steel material.
The finely crafted front panel and tabletop of the chassis, coupled with a unique humanized R-rounded design structure, can prevent operators from being accidentally injured by sharp edges.
The drainage system can meet the needs of various laboratories and enterprises for liquid exchange during production.
¡ñ Safety protection device:
1. Equipped with power-off protection function
2. Equipped with upper and lower temperature alarm setting function
3. Low water level protection and alarm function
4. Temperature runaway protection and alarm function
5. The heating device has low water level anti dry burning and explosion-proof functions
6. Multiple protection functions for compressor overheating and overcurrent
¡ñ Optional Function Guide
1. An optional extended external PT100 temperature sensor can be used to detect and control the temperature of the external system in real-time during external circulation.
2. Optional built-in 1-30 segment temperature control program or constant speed program temperature control, which can perform automatic program temperature control and display the real-time running status of the set temperature time program.
3. Optional RS232 or RS485 communication interface, easy to connect to the upper computer, with Modbus RTU communication protocol.
4. Optional design and installation of magnetic stirring system, which can directly perform magnetic stirring on liquid samples placed in beakers in the constant temperature bath working room, without the need for an external vertical mixer, reducing redundant operations.
5. Optional cooling coils built into the tank can quickly cool and control the exothermic reaction of the system.
6. Optional height adjustable immersion platform for easy adjustment of the liquid level of the immersed sample.
Comparison Table of Models and Parameters of XODCW Horizontal Low Temperature Constant Temperature Bath Series
model |
Temperature range (¡æ) |
Temperature uniformity (horizontal/vertical ¡æ) |
Inner groove size (mm)
(Length X Width X Slot Depth)
|
Total capacity (L) |
Work slot opening (mm2) |
Maximum cooling capacity (W) |
External circulation pump flow rate (L/min) |
XODCW-0506 |
-5~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
600 |
13 |
XODCW--0510 |
-5~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
900 |
13 |
XODCW--0515 |
-5~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
1500 |
13 |
XODCW--0520 |
-5~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
2000 |
13 |
XODCW--0530 |
-5~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
2700 |
13 |
XODCW--1006 |
-10~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
800 |
13 |
XODCW--1010 |
-10~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
1200 |
13 |
XODCW--1015 |
-10~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
1800 |
13 |
XODCW-1020 |
-10~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
2300 |
13 |
XODCW--1030 |
-10~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
3000 |
13 |
XODCW--2006 |
-20~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
1200 |
13 |
XODCW--2010 |
-20~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
1600 |
13 |
XODCW--2015 |
-20~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
2200 |
13 |
XODCW--2020 |
-20~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
2700 |
13 |
XODCW--2030 |
-20~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
3400 |
13 |
XODCW--3006 |
-30~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
1700 |
13 |
XODCW--3010 |
-30~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
2100 |
13 |
XODCW--3015 |
-30~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
2700 |
13 |
XODCW--3020 |
-30~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
3200 |
13 |
XODCW--3030 |
-30~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
3900 |
13 |
XODCW--4006 |
-40~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
2300 |
13 |
XODCW--4010 |
-40~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
2900 |
13 |
XODCW--4015 |
-40~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
3400 |
13 |
XODCW--4020 |
-40~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
3900 |
13 |
XODCW--4030 |
-40~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
4600 |
13 |
XODCW--5006 |
-50~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
2700 |
13 |
XODCW--5010 |
-50~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
3400 |
13 |
XODCW--5015 |
-50~100 |
¡À0.05 |
300¡Á250¡Á200j |
15 |
235¡Á160 |
3900 |
13 |
XODCW--5020 |
-50~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
4500 |
13 |
XODCW--5030 |
-50~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
5200 |
13 |
XODCW--6006 |
-60~100 |
¡À0.05 |
275¡Á195¡Á120 |
6 |
180¡Á140 |
3500 |
13 |
XODCW--6010 |
-60~100 |
¡À0.05 |
275¡Á195¡Á180 |
10 |
180¡Á140 |
4200 |
13 |
XODCW--6015 |
-60~100 |
¡À0.05 |
300¡Á250¡Á200 |
15 |
235¡Á160 |
4700 |
13 |
XODCW--6020 |
-60~100 |
¡À0.05 |
300¡Á250¡Á270 |
20 |
235¡Á160 |
5300 |
13 |
XODCW--6030 |
-60~100 |
¡À0.05 |
440¡Á325¡Á210 |
30 |
310¡Á280 |
6000 |
13 |
Manufacturer friendly reminder - When selecting a thermostatic bath, the following points should be considered:
Firstly, what is the range of temperature required for use? What medium is used as the temperature field carrier, and the medium used varies depending on the temperature range. The medium is generally selected according to the temperature range: ethylene glycol for water (low temperature), alcohol (ultra-low temperature), pure water (5-95 ¡æ), imported dimethyl silicone oil (suitable for high and low temperatures), etc.
2. Choose the appropriate capacity of the constant temperature bath based on your actual situation. The correct calculation method for effective working volume is: the opening size of the working bath plus the bath depth, in millimeters.
3. It is a requirement for the uniformity of the temperature field in the constant temperature bath. Our company divides it into two categories: 1. Ordinary type: ¡À 0.05 ¡æ. 2¡¢ Ultra high precision type: ¡À 0.0050 ¡æ (used as a calibration standard in the manufacturing of various thermometers, various types of thermistors, and other temperature measuring instruments)
4. Depending on the usage situation, what is the maximum head of the external circulation function.